top of page

Troubleshooting AMX

Error Number
Error 0785
Error Code / System Message
​
Error Question
What does error 0785 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0785: Config: Revision device configuration failure detected.
Possible Causes
Subsystem/category: Thermal. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Allow cooling when appropriate, verify airflow, and run generator thermal/70C diagnostics. Check the board or sensor identified by the code.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page