top of page

Troubleshooting AMX

Error Number
Error 0A81
Error Code / System Message
​
Error Question
What does error 0A81 mean on a GE Optima XR240amx?
Short Answer
Error 0A81: Internal power supplies out of specification.
Possible Causes
Subsystem/category: Misc. Manual recovery/error class: INHIBIT.
Troubleshooting Steps
Review the Class M / generator error log for related errors first. Restart only when appropriate, then verify generator software integrity, communications, power and cabling using the applicable SUIF diagnostics.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page