top of page
Troubleshooting AMX
Error Number
Error 0B86
Error Code / System Message
Error Question
What does error 0B86 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 0B86: mA: Filament ageing saturation detected when computed target reach command limit.
Possible Causes
Subsystem/category: mA Generation. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review related generator errors. Run mA Measurement and the applicable filament-driver tests; inspect tube/filament connections and use the specific error workflow before replacing parts.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx