Short Answer
Error 0B86: Filament aging saturation detected when computed target reaches command limit.
Possible Causes
Subsystem/category: mA. Manual recovery/error class: DEBUG.
Troubleshooting Steps
Review the generator error log for related errors. Run the applicable Filament Driver, Filament Drive Calibration and mA Measurement diagnostics. Inspect filament/tube connections and generator filament circuitry before replacing parts.
Parts Areas to Check
​
For Related Models
Optima XR240amx