top of page
Troubleshooting AMX
Error Number
Error 102
Error Code / System Message
ERROR 102 - Charge-current feedback low during high-current charging
Error Question
What does ERROR 102 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 102: Charge-current feedback low during high-current charging.
Possible Causes
Feedback was below expected high-current range.
Troubleshooting Steps
Check charger feedback, scale-select signal, charger-board connection and charge current.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page