top of page

Troubleshooting AMX

Error Number
Error 104
Error Code / System Message
ERROR 104 - Charge-current feedback low during trickle charge
Error Question
What does ERROR 104 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 104: Charge-current feedback low during trickle charge.
Possible Causes
Feedback was below expected trickle range.
Troubleshooting Steps
Check charger scale-select/feedback and actual trickle-current behavior.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page