top of page
Troubleshooting AMX
Error Number
Error 107
Error Code / System Message
ERROR 107 - Charge-clamp voltage exceeded at zero current
Error Question
What does ERROR 107 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 107: Charge-clamp voltage exceeded at zero current.
Possible Causes
Charge-clamp voltage exceeded the allowed condition.
Troubleshooting Steps
Check charger scale-select/feedback and charger-board output.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page