top of page

Troubleshooting AMX

Error Number
Error 1221245
Error Code / System Message
​
Error Question
What does error 1221245 mean on a GE Optima XR240amx?
Short Answer
Error 1221245: Access point boot failed.
Possible Causes
Subsystem/category: Wireless AP. Manual recovery/error class: WARNING.
Troubleshooting Steps
Wait for boot attempt to finish, reboot the system, then verify country configuration, AP Ethernet/power connections and AP operation.
Parts Areas to Check
​
For Related Models
Optima XR240amx