top of page

Troubleshooting AMX

Error Number
Error 1221217
Error Code / System Message
​
Error Question
What does error 1231217 mean on a GE Optima XR240amx?
Short Answer
Error 1231217: A non-compatible detector has been connected to the system.
Possible Causes
Subsystem/category: Detector. Manual recovery/error class: WARNING.
Troubleshooting Steps
Remove the non-compatible detector. Verify detector vintage, private-key applicability and compatibility before attempting registration again.
Parts Areas to Check
​
For Related Models
Optima XR240amx
bottom of page