Troubleshooting AMX

Error Number
Error 216
Error Code / System Message
ERROR 216 - Right drive feedback higher than command
Error Question
What does ERROR 216 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 216: Right drive feedback higher than command.
Possible Causes
Right drive-current feedback exceeded the CPU command.
Troubleshooting Steps
Check CPU-drive board connection, feedback buffers and drive-control electronics.
Parts Areas to Check
Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700