top of page

Troubleshooting AMX

Error Number
Error 217
Error Code / System Message
ERROR 217 - Right drive feedback lower than command
Error Question
What does ERROR 217 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 217: Right drive feedback lower than command.
Possible Causes
Right drive-current feedback was below the CPU command.
Troubleshooting Steps
Check CPU-drive board connection, feedback buffers and drive-control electronics.
Parts Areas to Check
Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page