top of page
Troubleshooting AMX
Error Number
Error 320
Error Code / System Message
ERROR 320 - 60 Hz inverter feedback missing at field-light command
Error Question
What does ERROR 320 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 320: 60 Hz inverter feedback missing at field-light command.
Possible Causes
60 Hz inverter feedback did not appear when field light was commanded.
Troubleshooting Steps
Check 60 Hz enable/relay/inverter signals and Generator Control 1 connections.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page