top of page
Troubleshooting AMX
Error Number
Error 426
Error Code / System Message
ERROR 426 - Autocalibration data fault
Error Question
What does ERROR 426 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 426: Autocalibration data fault.
Possible Causes
Auto-calibration data indicated a calibration error or potentially invalid information.
Troubleshooting Steps
Review calibration status and repeat the required generator calibration.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page