top of page

Troubleshooting AMX

Error Number
Error 432
Error Code / System Message
ERROR 432 - Filament-current demand feedback too high
Error Question
What does ERROR 432 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 432: Filament-current demand feedback too high.
Possible Causes
Filament current demand feedback exceeded expected value during prep.
Troubleshooting Steps
Check filament board/CPU connection, current feedback and tube filament connections.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page