top of page

Troubleshooting AMX

Error Number
Error 433
Error Code / System Message
ERROR 433 - Leakage-compensation command feedback high
Error Question
What does ERROR 433 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 433: Leakage-compensation command feedback high.
Possible Causes
Leakage-compensation feedback was too high during prep.
Troubleshooting Steps
Check leakage compensation feedback, filament board and kVp feedback.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page