top of page
Troubleshooting AMX
Error Number
Error 435
Error Code / System Message
ERROR 435 - Tap feedback does not match selected tap
Error Question
What does ERROR 435 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 435: Tap feedback does not match selected tap.
Possible Causes
Selected tap feedback was inconsistent during prep.
Troubleshooting Steps
Check tap-frequency/feedback signals, tap relay board and generator control.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page