top of page

Troubleshooting AMX

Error Number
Error 436
Error Code / System Message
ERROR 436 - X-RAY ON asserted during pre-exposure interlock check
Error Question
What does ERROR 436 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 436: X-RAY ON asserted during pre-exposure interlock check.
Possible Causes
X-RAY ON was active before exposure should begin.
Troubleshooting Steps
Check X-RAY ON status line, CPU critical-status input and generator-control sequencing.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page