top of page
Troubleshooting AMX
Error Number
Error 43A
Error Code / System Message
ERROR 43A - Rotor interlock feedback low during exposure interlock
Error Question
What does ERROR 43A mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 43A: Rotor interlock feedback low during exposure interlock.
Possible Causes
Rotor interlock feedback was not present when expected.
Troubleshooting Steps
Check rotor interlock signal, CPU critical-status input and generator-control feedback.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page