top of page

Troubleshooting AMX

Error Number
Error 44
Error Code / System Message
ERROR 44 - Battery voltage high during prep
Error Question
What does ERROR 44 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 44: Battery voltage high during prep.
Possible Causes
Battery voltage was higher than expected during prep.
Troubleshooting Steps
If just charged, remove surface charge by driving briefly; otherwise verify battery voltage/calibration.
Parts Areas to Check
Battery set; charger board; CPU board; voltage/current feedback wiring
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page