top of page
Troubleshooting AMX
Error Number
Error 444
Error Code / System Message
ERROR 444 - X-RAY PRESS status high in non-X-ray condition
Error Question
What does ERROR 444 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 444: X-RAY PRESS status high in non-X-ray condition.
Possible Causes
X-ray pressure/presence status remained high when it should be low.
Troubleshooting Steps
Check pressure/status switch, wiring and CPU critical-status input.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page