top of page
Troubleshooting AMX
Error Number
Error 445
Error Code / System Message
ERROR 445 - 60 Hz inverter status high before prep field light command
Error Question
What does ERROR 445 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 445: 60 Hz inverter status high before prep field light command.
Possible Causes
60 Hz inverter status was high at an unexpected time.
Troubleshooting Steps
Check 60 Hz inverter ON/relay signal and generator-controller path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page