top of page

Troubleshooting AMX

Error Number
Error 45A
Error Code / System Message
ERROR 45A - Leakage-compensation feedback low
Error Question
What does ERROR 45A mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45A: Leakage-compensation feedback low.
Possible Causes
Leakage-compensation feedback was lower than expected.
Troubleshooting Steps
Check leakage compensation feedback and filament-board/kVp feedback path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700