top of page
Troubleshooting AMX
Error Number
Error 45B
Error Code / System Message
ERROR 45B - Filament feedback low
Error Question
What does ERROR 45B mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45B: Filament feedback low.
Possible Causes
Filament feedback was lower than command.
Troubleshooting Steps
Check filament feedback, tube filament/cathode connections and filament board.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page