top of page

Troubleshooting AMX

Error Number
Error 45D
Error Code / System Message
ERROR 45D - Filament-short signal asserted during prep
Error Question
What does ERROR 45D mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45D: Filament-short signal asserted during prep.
Possible Causes
Filament short-detect signal was asserted.
Troubleshooting Steps
Check filament short-detect circuitry and associated generator wiring.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page