top of page

Troubleshooting AMX

Error Number
Error 45E
Error Code / System Message
ERROR 45E - Filament-short detect asserted after inverter start
Error Question
What does ERROR 45E mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45E: Filament-short detect asserted after inverter start.
Possible Causes
Filament short-detect status did not clear as expected.
Troubleshooting Steps
Check filament short-detect signal, tube/cable orientation and filament circuit.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700