top of page
Troubleshooting AMX
Error Number
Error 45F
Error Code / System Message
ERROR 45F - Filament-short signal remains active after filaments removed
Error Question
What does ERROR 45F mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 45F: Filament-short signal remains active after filaments removed.
Possible Causes
Short-detect remained active after filament loads were removed.
Troubleshooting Steps
Inspect high-voltage cable polarity/orientation and filament short-detect circuitry.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700