top of page

Troubleshooting AMX

Error Number
Error 461
Error Code / System Message
ERROR 461 - 60 Hz inverter OK signal low during exposure interlock
Error Question
What does ERROR 461 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 461: 60 Hz inverter OK signal low during exposure interlock.
Possible Causes
60 Hz inverter OK status was not present.
Troubleshooting Steps
Check inverter OK/status line, CPU input and inverter/relay circuitry.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page