top of page
Troubleshooting AMX
Error Number
Error 463
Error Code / System Message
ERROR 463 - kVp command feedback low during exposure interlock
Error Question
What does ERROR 463 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 463: kVp command feedback low during exposure interlock.
Possible Causes
kVp demand feedback was low during exposure interlock.
Troubleshooting Steps
Check CPU-generator connection, filament board and feedback path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page