top of page
Troubleshooting AMX
Error Number
Error 465
Error Code / System Message
ERROR 465 - Filament-current command feedback low during exposure interlock
Error Question
What does ERROR 465 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 465: Filament-current command feedback low during exposure interlock.
Possible Causes
Filament-current command feedback was below expected during exposure interlock.
Troubleshooting Steps
Check for excessive battery voltage drop, filament feedback and generator control.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700