top of page

Troubleshooting AMX

Error Number
Error 467
Error Code / System Message
ERROR 467 - Exposure determined to be short / stop condition
Error Question
What does ERROR 467 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 467: Exposure determined to be short / stop condition.
Possible Causes
Exposure was terminated by an early stop condition.
Troubleshooting Steps
Check exposure-stop logic, generator command path and associated errors.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700