top of page

Troubleshooting AMX

Error Number
Error 469
Error Code / System Message
ERROR 469 - X-RAY ON status not low during exposure shutdown
Error Question
What does ERROR 469 mean on a GE AMX-Series mobile X-ray?
Short Answer
ERROR 469: X-RAY ON status not low during exposure shutdown.
Possible Causes
X-RAY ON remained active when the exposure should have ended.
Troubleshooting Steps
Check X-RAY ON status, generator shutdown, kVp output and exposure-control path.
Parts Areas to Check
Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page