top of page
Troubleshooting AMX
Error Number
Error 8301072
Error Code / System Message
Error Question
What does error 8301072 mean on a GE Optima XR200amx or XR220amx?
Short Answer
Error 8301072: Battery pack 2 voltage over 50V – breaker to trip.
Possible Causes
Subsystem/category: System Battery. Manual recovery/error class: TRIP.
Troubleshooting Steps
Confirm whether the overvoltage follows pack 2 when pack positions are swapped; evaluate charging controls.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page