top of page

Troubleshooting AMX

Error Number
HIGH FIL LIMIT
Error Code / System Message
HIGH FIL LIMIT - High filament-current calibration limit exceeded
Error Question
What does HIGH FIL LIMIT mean on a GE AMX-Series mobile X-ray?
Short Answer
HIGH FIL LIMIT: High filament-current calibration limit exceeded.
Possible Causes
One or more filament-current calibration parameters exceeded the high limit.
Troubleshooting Steps
Verify filament current, mAs/kVp calibration and tube connections; repeat filament calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page