top of page
Troubleshooting AMX
Error Number
HNDL CAL ERR 1
Error Code / System Message
HNDL CAL ERR 1 - Drive-handle zero point out of range
Error Question
What does HNDL CAL ERR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
HNDL CAL ERR 1: Drive-handle zero point out of range.
Possible Causes
Drive handle zero-point value is outside expected range.
Troubleshooting Steps
Recalibrate drive handle and inspect Hall-effect transducers/supply.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page