top of page

Troubleshooting AMX

Error Number
HNDL CAL ERR 3
Error Code / System Message
HNDL CAL ERR 3 - Calculated reverse gain out of range
Error Question
What does HNDL CAL ERR 3 mean on a GE AMX-Series mobile X-ray?
Short Answer
HNDL CAL ERR 3: Calculated reverse gain out of range.
Possible Causes
Reverse handle-calibration gain is outside range.
Troubleshooting Steps
Recalibrate handle; inspect transducer output and CPU input.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page