top of page

Troubleshooting AMX

Error Number
KVP CAL ERROR 4
Error Code / System Message
KVP CAL ERROR 4 - kVp calibration parameter out of limits
Error Question
What does KVP CAL ERROR 4 mean on a GE AMX-Series mobile X-ray?
Short Answer
KVP CAL ERROR 4: kVp calibration parameter out of limits.
Possible Causes
One or more kVp calibration parameters exceeded allowed limits.
Troubleshooting Steps
Verify kVp values/meter setup and repeat kVp calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700