Troubleshooting AMX

Error Number
12 V debug-header value low or zero
Error Code / System Message
12V debug header
Error Question
What does low or zero 12 V on the Definium AMX 700 debug header indicate?
Short Answer
Very low 12 V suggests a current overload; 0 V can result from converter over-temperature, input over/under-voltage, converter failure or other supply faults.
Possible Causes
12 V output overload; converter over-temperature; input voltage fault; converter failure.
Troubleshooting Steps
Measure debug-header connection 2. If very low, isolate 12 V loads. If 0 V, check converter temperature/input conditions and the 12 V converter.
Parts Areas to Check
Digital Power Supply; 12 V loads
For Related Models
Definium AMX 700