top of page
Troubleshooting AMX
Error Number
MAS CAL ERROR 1
Error Code / System Message
MAS CAL ERROR 1 - mAs calibration frequency too high
Error Question
What does MAS CAL ERROR 1 mean on a GE AMX-Series mobile X-ray?
Short Answer
MAS CAL ERROR 1: mAs calibration frequency too high.
Possible Causes
Injected-current calibration produced a frequency above the allowed range.
Troubleshooting Steps
Check mAs test source, monitor connector and mA measuring circuit.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page