top of page

Troubleshooting AMX

Error Number
RAM READBACK ERR
Error Code / System Message
RAM READBACK ERR - RAM write/read verification failed
Error Question
What does RAM READBACK ERR mean on a GE AMX-Series mobile X-ray?
Short Answer
RAM READBACK ERR: RAM write/read verification failed.
Possible Causes
Data written to RAM did not read back correctly.
Troubleshooting Steps
Check RAM/data-bus connections and CPU board.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700