top of page

Troubleshooting AMX

Error Number
Error 3600134
Error Code / System Message
​
Error Question
What does system error 3600134 mean on XR200amx or XR220amx?
Short Answer
3600134 is the generic Generator Inhibit Non-Thermal Error.
Possible Causes
A non-thermal Djinn condition is inhibiting operation.
Troubleshooting Steps
Identify the detailed Djinn code in the Class M log and troubleshoot the indicated filament, rotor, DC-bus, kV, system, boot or communication subsystem.
Parts Areas to Check
​
For Related Models
Optima XR200amx, Optima XR220amx
bottom of page