top of page
Troubleshooting AMX
Error Number
TEST 02 FAILED
Error Code / System Message
TEST 02 FAILED - RAM battery test failed
Error Question
What does TEST 02 FAILED mean on a GE AMX-Series mobile X-ray?
Short Answer
TEST 02 FAILED: RAM battery test failed.
Possible Causes
Non-volatile RAM battery voltage/retention is below the expected condition.
Troubleshooting Steps
Check RAM battery, RAM circuitry and CPU board.
Parts Areas to Check
CPU board; memory/timer/A-D circuitry identified by the test
For Related Models
AMX-4, AMX-4+, Definium AMX 700
bottom of page