Troubleshooting AMX

Error Number
TEST 04 FAILED
Error Code / System Message
TEST 04 FAILED - Calibration data checksum failed
Error Question
What does TEST 04 FAILED mean on a GE AMX-Series mobile X-ray?
Short Answer
TEST 04 FAILED: Calibration data checksum failed.
Possible Causes
One or more calibration-data records do not match their stored checksum. Calculates the Check Sum for the Calibration Data Base and compares it with the stored value. Test fails if they are not equal. Entering Calibration will give a more specific error code for the check sum failure. Error code 84 Hex is recorded in the Error Code for a failure.
Troubleshooting Steps
Check Calibration and RAM. Identify the affected calibration, correct it, and rerun checksum/calibration.
Parts Areas to Check
CPU board; memory/timer/A-D circuitry identified by the test
For Related Models
AMX-4, AMX-4+, Definium AMX 700