top of page

Troubleshooting AMX

Error Number
1V5 LED
Error Code / System Message
1V5 - DS19 GREEN
Error Question
What does the 1V5 LED mean on the GE Definium AMX 700?
Short Answer
The FPGA core voltage is present.
Possible Causes
1.5 V FPGA core power present
Troubleshooting Steps
Use the LED state together with the system symptom. If the LED indicates an error, inhibit or temperature condition, perform the corresponding subsystem troubleshooting before replacing the board.
Parts Areas to Check
Beetle Interface Board
For Related Models
Definium AMX 700
bottom of page