top of page

Troubleshooting AMX

Error Number
Unit fails startup self-test
Error Code / System Message
TEST 00-07
Error Question
What does the AMX-Series power-up diagnostic check?
Short Answer
The startup diagnostic verifies the microcontroller, EPROM checksum, non-volatile RAM battery, external RAM, calibration checksum, watchdog timer, programmable interval timer, and A/D converter/power-supply circuitry.
Possible Causes
Failure of one of the startup hardware or firmware integrity tests.
Troubleshooting Steps
Observe the TEST number shown during power-up. Use TEST 00 through TEST 07 to isolate the failing subsystem, then follow the matching power-up error-code action.
Parts Areas to Check
CPU board; RAM; EPROM; timers; A/D converter; power supplies
For Related Models
AMX-4, AMX-4+, Definium AMX 700