top of page
Troubleshooting AMX
Error Number
Power control comm issues
Error Code / System Message
Error Question
What does the Brivo XR285amx Power Control Comm diagnostic test?
Short Answer
It checks low-level communication and control between the Spyder control board and other hardware controllers.
Possible Causes
Spyder or controller communication fault; cable or board issue.
Troubleshooting Steps
Run Power Control Comm in SUIF. If it fails, use the error log and System Sanity/communication results to isolate the controller, wiring, or board that is not communicating.
Parts Areas to Check
For Related Models
Brivo XR285amx
bottom of page