Short Answer
It checks the internal ±13 V, 48 V, and 3.3 V generator supplies while modules are off.
Possible Causes
Generator internal supply fault; 48 V LVLE2 issue.
Troubleshooting Steps
Run Power Supplies. For 48 V faults, verify LVLE2 output is 48 VDC ±2.5 V with the specified test setup and inspect the CAN/power connections.