top of page
Troubleshooting AMX
Error Number
TURN RATIO LIMIT
Error Code / System Message
TURN RATIO LIMIT - Turn-ratio calibration limit exceeded
Error Question
What does TURN RATIO LIMIT mean on a GE AMX-Series mobile X-ray?
Short Answer
TURN RATIO LIMIT: Turn-ratio calibration limit exceeded.
Possible Causes
Battery-voltage to kVp multiplication factor is outside range.
Troubleshooting Steps
Verify mAs/kVp calibration and transformer/tap relationships.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700