top of page

Troubleshooting AMX

Error Number
Prep/expose problem
Error Code / System Message
EXPOSR / X-RAY ON LEDs
Error Question
What should I check if the Definium AMX 700 will not prep or expose?
Short Answer
The Beetle functional test verifies an exposure, the audible tone and X-ray ON LEDs.
Possible Causes
Exposure inhibit; Beetle Board signal issue; generator/AMX exposure-circuit issue.
Troubleshooting Steps
Perform the approved Prep/Expose functional test. Verify the exposure tone and X-ray ON indicators. If the EXPOSR LED is lit, investigate the exposure-inhibit condition first.
Parts Areas to Check
Beetle Board; AMX exposure control; generator
For Related Models
Definium AMX 700