Short Answer
Several application errors are generated when X-RAY ON is asserted at the wrong time or is missing when expected.
Possible Causes
X-RAY ON signal wiring; status-port fault; generator control or inverter sequence problem.
Troubleshooting Steps
Check the X-RAY ON signal at the Critical Status port, verify the CPU connection and generator-control timing, and inspect the related relay/inverter command path.
Parts Areas to Check
Generator control board; CPU board; X-ray status line
For Related Models