top of page
Troubleshooting AMX
Error Number
Charging bin not recognize the FlashPad detector
Error Code / System Message
Error Question
What should I check when the XR220amx charging bin does not recognize the detector?
Short Answer
Detector Parked/bin-sense functions can be checked from SUIF; detector troubleshooting also distinguishes charging-bin sensing from table/wall-stand docking.
Possible Causes
Bin switch/sense issue; detector not seated; charging-bin power/connector problem.
Troubleshooting Steps
Confirm the detector is fully seated, run Detector Parked, observe charging/indicator status and compare another detector if available. Inspect bin contacts/connectors if the fault is isolated to the bin.
Parts Areas to Check
For Related Models
Optima XR200amx, Optima XR220amx